ホーム 格闘技 Summary of Impurities in NMR

Summary of Impurities in NMR

藤崎 健人 藤崎 健人 • 更新: • 公開: • 読了約 6 分
目次
n13063b923156

Impurity peaks are a concern in NMR measurements. Since there are almost no places where these peaks are summarized, I have recorded them here as a memorandum.

Plasticizers

Since they are dialkyl phthalate derivatives, they are troublesome substances that appear in the aromatic region.

How to identify: In 1H NMR spectra, they are characterized by two multiplets near deuterated chloroform at 7.4-7.8 ppm and a multiplet near 4 ppm.
How to remove: It is recommended to perform column chromatography carefully.

Reaction products of ether solvents

These sometimes mix in when THF or dioxane are reacted under strongly acidic conditions. 4-chlorobutan-1-ol is produced from THF.

How to identify: In 1H NMR spectra, two ABq signals can be seen near 3.5 ppm. Be careful, as this is sometimes seen when using THF.
How to remove: Since the boiling point is low at around 80°C, it can be removed by applying a vacuum and heating to about 40°C. It is easy.

Stabilizers

BHT (4-tert-butyl-2,6-dimethylphenol) contained in ether solvents.

How to identify: One singlet is seen in the aromatic region, and signals derived from the benzyl position near 2.0 ppm are seen for 6H. A phenol signal may also be visible.
How to remove: Because of its high polarity, it can be easily removed by a column.

Grease

Grease used for creating a vacuum can sometimes contaminate the sample. There are two types, silicone-based and alkyl-based, but since most people use silicone grease, impurities from silicone-based grease are often seen.
※ Recently, it has been found that it can be seen when using plastic columns often used in medium-pressure preparative column chromatography. It is thought that dimethyl silicone-based substances may be leaching from the column.

How to identify: A singlet is seen at 0.08 ppm.
How to remove: It dissolves in low-polarity solvents (especially hexane). If you can recrystallize using hexane as a poor solvent, do that. Alternatively, perform column chromatography using hexane as the developing solvent and add a polar solvent to elute the target compound.

同じカテゴリの記事

藤崎 健人

芸能取材歴10年のデスク。所属事務所の公式リリース、番組公式サイト、配信サービスの告知を一次情報として採用し、裏付けの取れない内容は記事にしていません。

おすすめの関連記事