I attended the Aural (Re)habilitation Course held at Singapore General Hospital on August 3-4, 2026, along with speech-language pathologists from the Cochlear Implant Center at Tokyo Medical Center. Focusing on the lecture content by instructor Dr. René Gifford (Hears for Hearing), I will summarize cochlear implant programming parameters as part of an 'Understanding Cochlear Implant Mapping' series.
The first installment is 'Threshold (T/THR) — How is the 'onset of hearing' determined?'
What parameter determines the Threshold (T/THR)?
The fact that the 'definition of T/THR itself' differs among the three companies (Advanced Bionics/Cochlear/MED-EL)
Why is an 'aided threshold of 20-30 dB HL' the goal? And why should you not make it better than that?
Why ECAP (NRT/NRI/ART) should not be used to determine T/THR
1. Definition: T/THR is the 'lower end' of EDR and IDR
Cochlear implants convert sounds picked up by the microphone into electrical stimulation and deliver them to the auditory nerve via electrodes in the cochlea. At this time, there is a relationship where:
Current that is too small → Nothing is felt
Just the right amount of current → Felt as sound
Current that is too large → Noisy, uncomfortable
The boundary between 'feeling nothing' and 'being able to feel' is T/THR.
And the range above this boundary, up to the point just before it becomes 'noisy,' is called the electrical dynamic range (EDR). T/THR determines the lower end of this EDR.
The definition of T/THR differs among the three companies.Advanced Bionics (T) : Psychophysical threshold (level at which 50% detection is possible)
Cochlear (T) : The minimum level that can be detected with 100% probability
MED-EL (THR): The maximum level that is not heard (= placed on the side of not being heard)
The THR for MED-EL has a philosophy opposite to the other two companies. It is set not at the level where "hearing begins," but at the level where "there is no hearing yet." In practice, it is standard to set the THR to 8% (or 0 qu) of the MCL.
Why does such a difference arise?This is because each company has a different design philosophy for signal processing. In particular, at MED-EL, a conversion curve called maplaw (which will be discussed later) works to push small sounds into a higher position within the electrical dynamic range, so even if the THR is set to a point where there is "no hearing," actual small sounds are sufficiently audible. In fact, if the THR is set too high, the user will feel a "hissing" background stimulation during silence.
2. "How small of a sound is mapped to T?"
T/THR is a setting for "current amount," but what is clinically important is "what dB SPL sound is mapped to T?" This also differs among the three companies.
Advanced Bionics: Depends on the IDR setting.
The formula is 65 dB SPL − IDR
・Programmable range of IDR: 20–80 dB / Recommended range: 60–80 dB
・Example: If IDR = 70 dB, then −5 dB SPL is mapped to TCochlear: 25 dB SPL
(can fluctuate due to ASC operation or manual changes to TSPL)MED-EL: 25 dB SPL
It is important that only Advanced Bionics is "dependent on IDR." With AB, even without touching the T level itself, just changing the IDR changes "how much small sound is picked up and delivered." This relationship will be covered in detail in the IDR article (③ Input Dynamic Range).
3. The target for verification is the "CI-aided threshold"
Setting T/THR is not the end. It is necessary to confirm that the settings are working as intended (= verification).
The verification target for T/THR is the CI-aided thresholds. In other words, we perform a hearing test in a sound field (speakers in a soundproof room) while the cochlear implant is worn, and draw an audiogram.
Goal is 20–30 dB HL(In the Hearts for Hearing protocol, the goal is 20–30 dB HL at 1 month post-activation)
Why is the aided threshold so important?Because audibility is a prerequisite for speech understanding.
In the field of audiology, a method of visualizing the relationship between aided thresholds and conversational speech using a diagram called a Count-the-Dots audiogram is used. We count how many dots of conversational level speech (60 dB SPL) fall above the aided threshold, and express this as the Speech Intelligibility Index (SII).
In Dr. Gifford's lecture, even with the same "cochlear implant in use"
Case with poor aided threshold → Aided SII ≒ 42%
Case with good aided threshold → Aided SII ≒ 95%
These two cases are being contrasted.This means that the same conversational sound reaches the user at only 40% in one case, while it reaches over 90% in the other.
This is also supported by clinical data.
Skinner et al. (1999): Setting T-levels above "100% detection" results in sound field thresholds of 30 dB HL or better.
Davidson et al. (2010): In 109 adolescent users who have used cochlear implants since childhood, the group using a new processor with aided detection thresholds of 30 dB HL or better performed best in speech understanding of soft sounds. Furthermore, increased access to soft sounds did not impair hearing in noise.
The latter point, that "hearing in noise did not worsen," is very important in practice. The concern that "if you try to pick up soft sounds, it will become harder to hear in noisy places" was not supported, at least in this study.
4. So, is the aided threshold better the lower it is?
―― The answer is No.This is the point most easily misunderstood.
Dr. Gifford poses the following question in his lecture:
"Why be satisfied with an aided detection threshold of 20–25 dB HL? Why not aim lower (better)?"
And the correct answer provided is as follows:
" Aided thresholds better than 20–25 dB HL carry the risk of picking up environmental noise and circuit noise, which interferes with speech understanding and subjective sound quality evaluation."
On the other hand, the following three misconceptions are also presented.
❌ "Cochlear implants cannot achieve thresholds better than 20–25 dB HL" → It is technically possible.
❌ "Hearing-impaired individuals cannot achieve aided thresholds in the normal hearing range" → It is not that they cannot be achieved.
❌ "There are no meaningful sounds below 20–25 dB HL" → The sounds themselves do exist.
In other words, it is not 20–25 dB HL because it "cannot" be better, but rather it is 20–25 dB HL because "if you try to pick up sounds below that, you end up amplifying noise that doesn't need to be picked up, which is actually disadvantageous."
5. By Manufacturer: How to set and verify T/THR
Advanced Bionics: Ensure a wide IDR / T is around 10% of M / Utilize SoftVoice.
Cochlear:
Adults: Measured or counted T
Children: Minimum Response Level (MRL) (VRA, CPA, BOA, CAEP)MED-EL: THR is 0 qu or 8% of MCL (= inaudible THR)
Adults: Confirm with live voice at Sweep/Volume 0%
Children: THR ≒ level where no CAEP response is elicited level
To add a note about AB, many users are programmed with T set to 0% or 10% of M. If T is not measured, a wide IDR acts as an insurance policy to ensure audibility of low-level sounds.
However, this should not be read as "measuring T is unnecessary." Baudhuin et al. (2012) compared programs with T set to "10% of M" versus programs set to "10 CU below the level judged as soft" in 11 children aged 7–17 using AB, and reported that detection thresholds for all six Ling sounds were significantly lower with the latter (setting based on actual measurement). Furthermore, the optimal IDR/Sensitivity combination varied for each subject.
6. Why ECAP is not used to determine T/THR
ECAP (Electrically Evoked Compound Action Potential) is an objective measurement included by manufacturers under the names NRI (AB), NRT (Cochlear), and ART (MED-EL). It has the major advantages of requiring no additional equipment, no sedation, and being measurable even in children who cannot sit still. It is said to be measurable in approximately 95% of CI users.
Even so, ECAP should not be the basis for determining T/THR. There are three reasons.
The level that elicits an ECAP threshold falls within the "audible dynamic range" to begin with. In other words, the ECAP threshold is not T, but somewhere above T.
ECAP is measured at an extremely low stimulation rate. AB is 32 pps, Cochlear is 80 pps, and MED-EL is 40 pps. On the other hand, daily use maps operate in the hundreds to thousands of pps range. The measurement conditions and usage conditions are too different.
ECAP is a poor predictor of both T/THR levels and profile shape. This has been repeatedly shown by Morita et al. (2003), Gordon et al. (2004), Jeon et al. (2010), and others.
In the context of troubleshooting, Dr. Gifford clearly stated when examining "children with poor clarity" to "suspect a swooping MAP (U-shaped or V-shaped distorted profile) or an ECAP-based map".
This does not mean "do not use" ECAP.It is useful for recording baselines and checking electrode function, and eCAP is measured "for baseline purposes" in the Hearts for Hearing activation protocol as well. The problem is converting ECAP values directly into T/C to build a map.
7. Signs that users and their families can notice
When T/THR is too low or too high, it manifests in daily life in the following ways.
Signs that sounds may be too quiet (under-mapping)Taking off the processor or hearing aid
Wants to increase the volume
Strongly relies on visual cues such as lip movements
Does not respond or turn around when called by name
(For children) No change in vocalization when wearing the processor versus when it is removed
Speaks in a loud voice because their own voice sounds quiet to them
Frequently removes the processor. Short usage time
Keeps the volume turned down (especially at the minimum setting)
(For children) Startled by or cries at loud sounds
Involuntary blinking
Speaks in a quiet voice
Hearing actually worsens when the presentation level is high
The logic for decision-making presented by Dr. Gifford is simple.
Sound field threshold is poor (high) → Increase T/THR
Sound field threshold is good but sounds are quiet → Adjust C/M/MCL overall rather than T/THR
Sound field threshold is less than 20 dB HL at multiple frequencies (= too good) → Decrease T/THR
Sound field threshold is 20–25 dB HL but sounds are too loud → Decrease C/M/MCL overall
In other words, 'Hard to hear = increase T' is not necessarily true. First, measure the sound field threshold to determine whether the issue is on the lower end or the upper end. If you skip this step, you may end up just increasing background noise by tinkering with T.
8. Summary
T/THR is the anchor for the "start of hearing" and defines the lower end of the EDR and IDR.
The definition itself differs among the three companies. AB is 50% detection, Cochlear is 100% detection, and MED-EL is the "maximum level at which no sound is heard."
The loudness of the sound mapped to T is around 25 dB SPL for Cochlear/MED-EL, and for AB, it depends on the IDR (65 - IDR).
The target for verification is the CI aided threshold. The goal is 20–30 dB HL.
Do not make it better than that. This is because it picks up environmental noise and circuit noise, which actually impairs speech understanding and sound quality.
Do not use ECAP to determine T/THR. The measurement conditions are too different, and the prediction is poor.
If possible, determine T through behavioral measurement. Default values (such as 10% of M) are a safety net, not the optimal solution.
Next time, we will cover the Upper Stimulation Level (C/M/MCL), which is the other anchor of the EDR. Here, eSRT (electrically evoked stapedial reflex threshold) will take center stage.